0

    Analysis Examples

    Shim-pack FC-ODS separates components by hydrophobic interaction like other ODS columns. It is possible to change to Shim-pack FC-ODS from other ODS columns under the same analytical conditions. On the other hand, hydrophilic interaction (hydrogen bond, coordination bond) has been restricted to a minimum, which ensures significant efficiency when analyzing basic compounds. In addition, Shim-pack FC-ODS has higher steric selectivity (capability to recognize the difference of steric structures), making it possible to separate some components that are difficult to retain in other ODS columns.
    Shim-pack FC-ODS is available in three lengths to suit analysis objectives.

    ■ 30mm

    Recommended for use in high-throughput analysis of samples that do not have a complex matrix.

    ■ 75mm

    Recommended for shortening the analysis time to that of a 150 mm column. Because Shim-pack FC-ODS retains a similar number of theoretical plates as a 150 mm column, it is possible to obtain the same result within about half of the time without changing the conditions. (In the case of gradient analysis, it is necessary to change the concentration.)

    ■ 150mm

    Recommended for analyzing samples that are difficult to be retained in other 150 mm ODS columns.

    High-Throughput Analysis

    10 times repeated analysis in 6.5 minutes is possible.

    Shortens the analysis time using a conventional system

    High Speed Analysis of 2,4-DNPH derivatized aldehydes / ketones

    Chromatograms of Environmental Air (A)
    and Indoor Air at Laboratory (B)

    Specifications

      Shim-pack FC-ODS
     Particle Size  3 µm
     Pore Size  12 nm
     Surface Area  315 m2/g
     Carbon Loading  18%
     Pressure Tolerance  20 MPa
     Pore Volume  1 mL/g
     End-capping  Yes
     Bonding Type  Monomeric
     pH range  1.5-9
     USP Code  L1